LV SERIES VISCOMETERS & RHEOMETERS - (ENTIPOISEC
mpa.s)
Cone Spindle* |
LVT |
LVDV-I+ |
LVDV-II+* |
LVDV-III+ |
Shear
Rate*** (sec-1) |
Sample
Volume (mL) |
CPE-40 |
CP-40 |
0.5 -1,023 |
0.3 - 1,023 |
0.2 - 3,070 |
0.1 - 3,070 |
7.50N |
0.5 |
CPE-41 |
CP-41 |
1.9 - 3,837 |
1.2 - 3,837 |
0.6 - 11,510 |
0.5 - 11,510 |
2.00N |
2.0 |
CPE-42 |
CP-42 |
1.0- 2,000 |
0.6 - 2,000 |
0.3 - 6,000 |
0.2 - 6,000 |
3.84N |
1.0 |
CPE-51 |
CP-51 |
8.0 - 16,000 |
4.9 - 16,000 |
2.4 - 48,000 |
1.9 - 48,000 |
3.84N |
0.5 |
CPE-52 |
CP-52 |
15.4 - 30,720 |
9.3 - 30,720 |
4.6 - 92,160 |
3.7 - 92,160 |
2.00N |
0.5 |
RV SERIES VISCOMETERS & RHEOMETERS -
(ENTIPOISEC mpa.s)
Cone Spindle* |
RVT |
RVDV-I+ |
RVDV-II+* |
RVDV-III+ |
Shear
Rate*** (sec-1) |
Sample Volume (mL) |
CPE-40 |
CP-40 |
3.3 - 6,540 |
3.3 - 10,900 |
1.6 - 32,700 |
1.3 - 32,700 |
7.50N |
0.5 |
CPE-41 |
CP-41 |
12.3 - 24,560 |
12.3 - 40,933 |
6.1 - 122,800 |
4.9 - 122,800 |
2.00N |
2.0 |
CPE-42 |
CP-42 |
6.4 - 12,800 |
6.4 - 21,300 |
3.2 - 64,000 |
2.6 - 64,000 |
3.84N |
1.0 |
CPE-51 |
CP-51 |
51.2 - 102,400 |
51.2 - 170,666 |
25.6 - 512,000 |
20.5 - 512,000 |
3.84N |
0.5 |
CPE-52 |
CP-52 |
98.3 - 30,720 |
98.3 - 327,666 |
49.2 - 983,000 |
39.3 - 983,000 |
2.00N |
0.5 |
HA SERIES VISCOMETERS & RHEOMETERS - (ENTIPOISEC
mpa.s)
Cone
Spindle* |
HAT |
HADV-I+ |
HADV-II+* |
HADV-III+ |
Shear Rate*** (sec-1) |
Sample Volume (mL) |
CPE-40 |
CP-40 |
6.6 - 13,080 |
6.6 - 21,800 |
3.3 - 65,400 |
2.6 - 65,400 |
7.50N |
0.5 |
CPE-41 |
CP-41 |
24.6 - 49,120 |
24.6 - 81,866 |
12.3 - 245,600 |
9.8 - 245,600 |
2.00N |
2.0 |
CPE-42 |
CP-42 |
12.8 - 25,600 |
12.8 - 42,600 |
6.4 - 128,000 |
5.1 - 128,000 |
3.84N |
1.0 |
CPE-51 |
CP-51 |
102.4- 204,800 |
102.4 - 341,333 |
51.2 - 1,024,000 |
41.0 - 1,024,000 |
3.84N |
0.5 |
CPE-52 |
CP-52 |
196.6 - 393,200 |
196.6 - 655,333 |
98.3 - 1,966,000 |
78.6 - 1,966,000 |
2.00N |
0.5 |
HB SERIES VISCOMETERS & RHEOMETERS
Cone
Spindle* |
HBT |
HBDV-I+ |
HBDV-II+* |
HBDV-III+ |
Shear Rate*** (sec-1) |
Sample Volume (mL) |
CPE-40 |
CP-40 |
26.2 - 52,400 |
26.2 - 87,200 |
13.1 - 261,600 |
10.5 - 261,600 |
7.50N |
0.5 |
CPE-41 |
CP-41 |
98.3 - 196,600 |
98.3 - 327,466 |
49.1 - 982,400 |
39.3 - 982,400 |
2.00N |
2.0 |
CPE-42 |
CP-42 |
51.2 - 102,400 |
51.2 - 170,666 |
25.6 - 512,000 |
20.5 - 512,000 |
3.84N |
1.0 |
CPE-51 |
CP-51 |
409.6 - 819,200 |
409.6 - 1,365,333 |
204.8 - 4,096,000 |
163.8 - 4,096,000 |
3.84N |
0.5 |
CPE-52 |
CP-52 |
786.4 - 1,572,860 |
786.4 - 2,621,333 |
393.2 - 7,864,000 |
314.6 - 7,864,000 |
2.00N |
0.5 |
*CPE-refers to electronic gap adjustment, while CP
- refers to manual gap adjustment. **Maximum viscosity range shown is
at 0.1 rpm. ***N=rpm Example:RV Viscometer with CPE-42 spindle at 10 rpm
has shear rate 3.84 ¥¹ 10 = 38.4 sec -1
Various cone spindles are available for use with Wells-Brookfield Cone/Plate
instruments.One cone spindle is provided with the instru-ment and is calibrated
for use with the sample cup. Additional cone spindles may be purchased and will
be calibrated for use with the same sample cup. Cone spindle dimen-sions are
shown in the table.
Cone Spindle Dimensions
CONE SPINDLE |
ANGLE(deg.) |
RADIUS(cm) |
CP-40 or CPE-40 |
0.8 |
2.4 |
CP-41 or CPE-41 |
3.0 |
2.4 |
CP-42 or CPE-42 |
1.565 |
2.4 |
CP-51 or CPE-51 |
1.565 |
1.2 |
CP-52 or CPE-52 |
3.0 |
1.2 |
Electronic Gap Setting Technology allows fast set-up time and minimizes
the possibility of "gap" adjustment errors. A bright LED lights up when
a "hit point" is achieved. A simple turn of the micrometer adjustment
ring and the "gap" is set, reducing the time required associated
with manual detection methods.
Brookfield's Electronic Gap Setting Technology
is a standard feature on our complete
|