Parameters
|
Range
|
Resolution
|
Accuracy
|
C401
|
C301
|
C201
|
C105
|
C104
|
C103
|
C102
|
C101
|
Chlorine Free
and Total
|
0-1.99ppm
|
0.01ppm
|
±0.02ppm
|
¡Ü
|
¡Ü
|
¡Ü
|
|
|
|
|
|
2.0-6.0ppm
|
0.1ppm
|
±0.2ppm
|
Cyanuric
Acid
|
5-90ppm
|
1ppm
|
±4ppm
|
¡Ü
|
|
|
|
|
|
¡Ü
|
|
pH
|
5.9-8.2pH
|
0.1pH
|
±0.1pH
|
¡Ü
|
¡Ü
|
|
|
|
|
|
¡Ü
|
Chlorine
Dioxide
|
0-3.79ppm
|
0.01ppm
|
±0.02ppm
|
|
|
|
|
|
¡Ü
|
|
|
3.8-11.4ppm
|
0.1ppm
|
±0.2ppm
|
Bromine
|
0-4.9ppm
|
0.02ppm
|
±0.03ppm
|
|
|
|
|
¡Ü
|
|
|
|
4.5-13.5ppm
|
0.2ppm
|
±0.3ppm
|
Ozone
|
0-1.39ppm
|
0.01ppm
|
±0.02ppm
|
|
|
|
¡Ü
|
|
|
|
|
1.4-4.1
|
0.1ppm
|
±0.2ppm
|
Measurement
Method
|
Photometric
|
Light
Soure
|
Light
Emitting Diode(LED)
|
Wavelength
|
525nm
|
Detector
|
Silicon
photodiode
|
Absorbance
Range
|
0
-2.5 Abs
|
Photometric
Precision
|
+/-0.0015Als
|
Calibration
Points
|
User
selectalble, 1 point per colorimetric test
|
Display
|
4-digit
14-segments customised liquid crystal display
with annunciators
|
Sample
Vials
|
Borosilicate
glass with screw caps, fill line and indexing
mark
|
Sample
Required
|
10ml(0.33oz)
|
Operating
Temperature Range
|
0-50¡É
|
Sample
Temperature Range
|
0-50¡É
|
Operating
Humidity Range
|
0-90% RH
non-condensing at 30¡É
|
Power
Supply
|
4 AAA alkaline
bateries
|
Battery
Life
|
>
3000 tests
|
Electromagntic
Compliance(EMC)
|
Emitted
lnterference-EN 61326 lmmunity to interference-EN
61326
|
Enclosure
Rating
|
IP67
|
Insulation
Rating
|
Pollution
degree2
|
Weight
|
Instrument
:
200g Instrument carrying case : 1.25kg
|
Dimensions(L×W×H)
|
Instrument
: 68×155×46mm Carrying case
: 160×350×120mm
|