BM-9,
BM-7, BM-FAST |
1. °¢ ÃøÁ¤ ´É·ÂÀº
0.1,0.2,1,2µµ(´Ü BM-9´Â 0.2,1,2µµ) 2. °³°³ÀÎÀÇ ¿ÀÂ÷¸¦ Á¦°ÅÇÏ°í Ç¥ÁØ »öµµ Sample ·ÎºÎÅÍ
Á¶Á¤À» ½±°Ô ¸¸µë. 3. Àü¿øÀÌ ³ª°¡´õ¶óµµ dataÁ¶Á¤ÀÌ ±â¾ï µÇ¾úÀ½. 4. ÃøÁ¤ÁöÁ¡³»¿¡ÀÇ
°¨µµÀÇ °ÅÄ¥±â¸¦
ÃÖ¼ÒÇÑ Á¦ÇѵǰÔ
¼³°è¸¦ ÇÑÁ¦Ç°À¸·Î
7¼¼±×¸àÆ®ÀÇ LED³ª
Çü±¤ Ç¥½ÃÆÇÀÇ
Æò±ÕÀûÀÎ ÈÖµµ¸¦
ÃøÁ¤ ÇÒ ¼ö ÀÖÀ½. 5. Ç¥ÁØ
ÀÎÅÍÆäÀ̽º(GP-IB ȤÀº
RS-232C:±¸ÀԽÿ¡ ÁöÁ¤¿äÇÔ)°¡ ³»ÀåµÇ¾î ÀÖ¾î°í ÀÚµ¿ ÃøÁ¤ÀÌ °¡´É Çϸç
in-line¿¡ ÀÇÇÑ
È¿°ú¸¦ ¹ßÈÖ ÃøÁ¤. 6. ¿©·¯°¡ÁöÀÇ ÃøÁ¤°è»ê°ú Ç¥½Ã ±â´ÉÀÌ
°¡´ÉÇÔ. 7. »ö,ÈÖµµ,»ö¿Âµµ´Â ¹°·ÐÀÌ°í »ö»ó,¹à±â,¼±¸íµµ¿¡ µû¸¥ CIELAB, CIELUV,»öÂ÷ ±×¸®°í Ç¥ÁØ°ª¿¡ ´ëÇÑ Â÷À̸¦ °è»êÇÏ°í ¼ø°£
µð½ºÇ÷¹ÀÌ°¡ µÊ. 8. X.Y.Z¿¡ µû¸¥ 3ä³ÎÀÇ
analog output´Â recorder, oscilloscopeµî¿¡ ÀÇÇÑ ±â·Ï,°üÂû¿¡
È¿°ú ¹ßÈÖ. 9. ¼±ÃøÁ¤¿¡ ÀÖ¾î¼ È¿°úÀûÀÓ. 10. Àü¿ø: AC adapter(Ç¥ÁØ ¾×¼¼¼¸®), External
battery(Ưº° ¾×¼¼¼¸®) 11. MB-7 FAST´Â °í¼Ó ÃøÁ¤ ¸ðµ¨ÀÓ.ÃøÁ¤½Ã°£ÀÌ
Á¹·¡ÀÇ 1/4(¾à 0.5ÃÊ)¿¡
ÃøÁ¤.ÆÛ½º³¯ ÄÄÇ»Å͸¦
»ç¿ë ÇÏ´õ¶óµµ
4¹è¼ÓÀÌ µÊ BM-7
ÇÑ±Û ¸Å´º¾ó º¸±â | |
|
|
MODEL |
BM-9 |
BM-7 |
BM-7 FAST |
ÃøÁ¤°¢ |
°íÁ¤(0.2,1,2µµ) |
°¡º¯(
0.1,0.2,1,2µµ)
|
ÈÖµµÇ¥½Ã¹üÀ§ (ÃøÁ¤°¢ 2° |
2µµ: 0.01-199,900 cd/§³ 0.1µµ: 0.1~1,999,000 cd/§³ 0.2µµ: 1-19,990,000 cd/§³ |
0.01~12.000.000(0.1°) cd/§³ |
ÃÖ¼ÒÃøÁ¤Áö¸§ |
0.95mmØ(ÃøÁ¤°ª0.2°) |
0.49mmØ(ÃøÁ¤°ª0.1°) |
Ãø Á¤ ½Ã °£ |
±4% of rdg 1 digit |
»öµµÃøÁ¤,»ö¿ÂµµÃøÁ¤ |
¾à 3ÃÊ |
¾à 2ÃÊ |
¾à 0.5ÃÊ |
»ö µµ Á¤ µµ |
x |
o
|
o
|
µðÁöÅÐÃâ·Â |
RS-232C |
RS-232C/GP-1B(±¸ÀԽà ÁöÁ¤) |
¾Æ³ª·Î±×Ãâ·Â |
o |
o
|
o
|
»ç ¿ë Àü ¿ø
|
9V°ÇÀüÁö(JIS 6F22)1°³ |
Àü¿ë
AC¾Æ´äÅÍ
|
¹«°Ô
|
¾à 0.45kg |
¾à 3kg |
| |
4.1 Ç¥Áر¸¼º
(1) BM-7 º»Ã¼ ------------------------ 1
(2) AC ¾Æ´äÅÍ ------------------------ 1
(3) ´ë¹°·»ÁîCAP ---------------------- 1
(4) Á¢¾È·»ÁîCAP ---------------------- 1
(5) CASE ---------------------------- 1
(6) ¾Æ³¯·Î±× Ãâ·Â Ç÷¯±× --------------- 1
(7) Ãë±Þ¼³¸í¼ ------------------------ 1
4.2 Ưº°ºÎ¼ÓÇ° (Option)
(1) AC ¾Æ´äÅÍ AD-10860
(2) AC ¾Æ´äÅÍ AD-12860
(3) AC ¾Æ´äÅÍ AD-22860
(4) AC ¾Æ´äÅÍ AD-24860
(5) BM-7¿ë Ãø»ö ÇÁ·Î±×·¥
(6) »ï°¢ ¥²Çü
(7) ¹Ìµ¿´ë S-3
(8) Ç¥Áعé»öÆÇ WS-2
(9) Attachment Lens AL-6
(10) Mesh Filter MF-100
(11) Çö¹Ì°æ ´ë¹°·»Áî ¾Æ´äÅÍ AL-4
(12) ITV ¾Æ´äÅÍ
(13) ¿ÜºÎ¹åµ¥¸®
(14) ¿ÜºÎ¹åµ¥¸® ÃæÀü¿ë AC ¾Æ´äÅÍ AD-1018
(15) ¿ÜºÎ¹åµ¥¸® ÃæÀü¿ë AC ¾Æ´äÅÍ AD-1018B
(16) ºÐ±¤°¨µµÃøÁ¤ Data
|
***»ó¼¼
Specifications ´õ º¸±â
|
Specifications |
BM-9 |
Detector Model |
BM-920D |
BM-910D |
BM-902D |
Measuring
field |
2" |
1" |
0.2"
|
Optical
system |
Object lens f = 36mm
F2.5 |
Viewing
field |
5" |
Measurement
distance |
350mm ~ ¡Ä |
Measurement
diamenter
(uits :mmØ) |
Measuring field |
Measurement distance
(m) |
0.35 |
0.4 |
0.6 |
0.8 |
1 |
3 |
5 |
2¢ª |
9.5 |
11.2 |
18.2 |
25.3 |
32.3 |
102 |
173 |
1¢ª |
4.7 |
5.6 |
9.1 |
12.7 |
16.2 |
51.1 |
86.1 |
0.2¢ª |
0.95 |
1.12 |
1.82 |
2.53 |
3.23 |
10.2 |
17.3 | |
|
*Differs
somewhat according the finishing precision of the aperture mirror. * Measurement
distance here is the distance from the tip of the attachment lens melal
piece. |
Minimum measurement
diamenter |
9.5mm dia. <1.2mm dia
when using AL-8 (optional)> |
4.8mm dia. <0.6mm
dia. when using AL-8 (optional)> |
0.95mm dia. <0.12mm
dial.when using AL-8 (optional)> |
Display |
4-digit LCD |
Photo
cell |
Silicone photodiode |
Spectral
sensitivity characteristics |
Whitin 8% (deviation from the relative
luminous efficiency)*JISC 1609~1993 |
Measurement range |
0.01~199,900cd/§³ |
0.1~1,999,000cd/§³ (10 tims of displayed
value is luminance value) |
1~19,990,000cd/§³ (100 times of displayed
value is luminance value) |
Auto 5-step range |
Precision |
±4% of rdg. ±1 digit |
±4% of rdg. ±1
digit |
±4% of
rdg. ±1 digit |
(standard light source A.23' ±3¡É, auto
range, 1 cd/§³ or above) |
(standard light source A.23' ±3¡É, auto
range, 10 cd/§³ or above) |
(standard light source A.23' ±3¡É, auto
range, 100 cd/§³ or above) |
Temperature
properties |
Within ±3%(0~40¡É, 23¡É as
standard) |
Humidity
properties |
Within ±3% (85% R.H. or lower, 60% R.H. as
standard) |
Analog signal
output |
0~3 Vmax 1mv/1 digit Response speed at
time of analog output: 1~30 ms at FAST |
RS output
conditions |
baud rate:2400 BPS Data length : 7 Parity
: ODD Stop bit : 1 |
Power
supply |
One 9V battery (6F22)<Operationg time
using continuously/approx. 13 hours when not using RS-232C, 5 hours when
using RS-232C> |
Operating
conditions |
Temperature: 0¡É~40¡É Humidity : 85% R.H. or
lower |
External
dimensions |
Approx. 190(L) x 105(W) x 56.5(H) mm /
Display unit : Approx. 131(L) x 73.5(W) x 33(H) mm / Detecor :
Approx. 75.5(L) x 105(W) x 56.5(H) mm |
Weight |
Display unit: Approx. 200g(including
battery) / Detector : Approx. 250g |
Specifications |
BM-7 / BM-7 Fast
|
Measuring field |
Variable Aperture Selections of 2¢ª,
1¢ª,0.2¢ª,0.1¢ª |
View finder field |
5¢ª |
Optical system |
Objective lens : f=80mm, F2.5; Eyepiece
lens adjustment range : ±5 diopter |
Photo cell |
3-Element Silicon Photodiode (X©ü, Y,
Z) |
Megasuring functions |
x,y,L:(x,y); Chromaticity coordinates,(L);
Luminance ±¡â |
u,v,L:(u,v);Chromaticity coordinates,
L;Luminance ±¡â |
XYZ:Tristimulus values (X,Y,Z)
±¡â |
Tc; color temperature and duv; deviation
±¡â |
CIE 1976 L * a*b ¡â Eab *
±¡â |
CIE 1976 L * u*v ¡â Euv *
±¡â |
* ±¡â means difference between standard
value and measured value. |
For example, ¡âx,¡ây, ¡âL will be given in x,
y, L setting |
Display
intervals |
BM-7(approx.2sec), BM-7 Fast (approx. 0.5
sec) |
Luminance
Unit |
Candela per square meter
(cdRange),Foot-Lambarts(fL)selectable |
Luminance masuring range (unit:cd/§³) |
Field |
2¢ª |
1¢ª |
0.2¢ª |
0.1¢ª |
Range1 |
0.01~30 |
0.04~120 |
1~3,000 |
4~12,000 |
Range2 |
0.03~90 |
0.12~360 |
3~9,000 |
12~36,000 |
Range3 |
0.1~300 |
0.4~1,200 |
10~30,000 |
40~120,000 |
Range4 |
1~3,000 |
4~12,000 |
100~300,000 |
400~1,200,000 |
Range5 |
10~30,000 |
40~120,000 |
1,000~3,000,000 |
4,000~12,000,000 |
Response time (unit:ms) |
FAST |
Range1 |
Range2 |
Range3 |
Range4 |
Range5 |
30 |
15 |
5 |
3 |
1 |
Luminance accuracy |
Luminane: ±4% of display
value |
(For light source A 23¡É measuring field
2¢ª, 1cd/§³ or above, autorange) |
Chromaticity: (x.y):within ±0.002 for
standard source A, |
within ±0.01 (O-55, Y-48, T-44, A-73B,
IRA-05) |
within ±0.03(R-61, B-46, V-44,
G-54) |
(Standard light source A + color filiter)
|
(23¡É measuring field 2¢ª, 1cd/§³ or above,
autorange) |
Caligration standard |
Select either the TOPCON calibration
(standard light source A, 23¡É ±3¢ª below 65% RH) or the user's
calibration standard |
Display |
LCD 20 digits x4 lines with illumination
function |
Output |
Digital |
GP-IB or RS-232C interface is selected
at the time of purchase (*external computer control is
possible) |
Anglog |
Anglog voltage output proportioned to
X2, Y and Z Response speed varies from 1 to 30 ms depending on the
range used |
Power
supply |
AC adapter (standard
accessory) External battery (special sccessory) |
Power
consumption |
8.5 VDC, approx.
450mA |
Operationg requirements |
Temperature : 0¡É to 40¡É , Humidity : below
85% RH |
External
dimensions |
325(L) x 120(W) x
165(H)mm |
Weight |
3kg (main unit
only)* |
Range |
Auto/Manual 5 steps |
°èÃø±âÀüü¸Þ´º welcome to sjscale@naver.com
268 Àú¿ï ¹× °èÃø±â±âÀÇ ¸ðµç
°Í(½ÅÁ¤Á¤¹Ð±â°è)
(Tel:055-237-5300~2, FaX:055-237-5007)
|