SPECIFICATIONS |
Phoenix-986AA |
Optic
System |
Wavelength Range |
190nm ~ 900nm |
Wavelength Repeatability |
0.2nm |
Wavelength Accuracy |
±0.3nm |
Monochromater |
Czerny-Turner Configuration |
Resolution |
over
0.3nm |
Spectral Bandwidth |
0.1nm, 0.2nm, 0.5nm, 1.0nm, 2.0nm(5steps with automatic
changeover) |
Flame
Analysis |
Sensitivity |
Cu
5-ug/ml > 0.850Abs |
Burner Head |
Titanium alloy burner |
Nebulizer |
High-efficiency glass nebulizer |
Position Adjustment |
Automatic changeover of flame and furnace |
Safety |
Automatic ignition of mixing air-acetylene gas with protection control
Automatic setting of optimum height for flame burner |
Graphite
Furnace Analysis |
Character Value(Cd) |
0.8Pg |
Temperature Range |
Ambient - 2650¡É |
Heating Program |
Up
to 9 steps with choices of ramp, ladder & full-power
Heating |
Heating |
Voltage feedback control when drying & ashing ; Optical temperature
control when atomizing |
Background
Correction |
Deuterium Lamp Background Correction |
1.0Abs |
Self-Reversal Background Correction |
3.0Abs |
Data
Processing |
Analytical method |
Flame, Graphite furnace & hydride |
Determination Method |
Calibration curves(1~2 order), standard addition method |
Repetitions |
1~20
with calculations of average value, SD and RSD |
Mainframe |
Light Source |
8
lamps, 2 lamps simultaneously lit(with one preheat) |
Power Supply |
110V/60Hz or 220V/50Hz three-phase AC 200W(mainframe), 5000W(graphite
furnace) |
Size |
Mainframe 110cm x 50cm x 45cm x 165 Ibs Graphite furnace 50cm x 50cm
x 45cm x 110 Ibs |