Àü±â/ÀüÀÚ°èÃø±âDigital Earth TesterTES-1700

 

 

TES-1700

 

¡ÜCapable of measuring earth voltage.
¡ÜData hold, MAX/MIN function.
¡ÜAuto power off function
¡Ü2mA measuring current permits earth resistance tests without tripping  earth leakage current breakers in
   the circuit under test.
¡ÜAutomatic warning when resistance of auxiliary earth spike is in excess of tolerance.
¡ÜDesigned to meet IEC1010-1 safety standard.

  

ELECTRICAL SPECIFICATIONS :

* Measuring Range and Accuracy (at 23¡É and 45~75%RH)

Function

Measuring range

Accuracy

Overload Protection

Earth Voltage

0 ~ 399.9V (50Hz, 60Hz)

±(1%rdg+4dgts)

400Vrms/1 minutes

Earth Resistance

0 ~ 39.99§Ù

±(2%rdg+0.2§Ù)

250Vrms/10 seconds

0 ~ 399.9§Ù

±(2%rdg+3dgts)

0 ~ 3999§Ù

 

* Current Consumption (typical Value with battery voltage of 12V)

Range

ON

40§Ù/ 400§Ù/4000§Ù

58mA

Earth V

22mA

 

 GENERAL SPECIFICATIONS :

  • Safety Standard : IEC1010-1 overvoltage category III , IEC 61557
  • Display : 3-3/4 digit liquid crystal display with a maximum reading of 3999
  • Insulation Resistance : More than 5M§Ù between the circuit and the housing case
                                     when measured with 500VDC
  • Withstand Voltage : 3700CV AC for 1 minute between the circuit and the housing case
  • Timer (Auto power off) : about 3 minutes
  • Dimensions : 190(L) × 140(W) × 77(H)mm
  • Weight : Approx. 800(with battery)
  • Power Source : Eight 1.5V 01t AA Size Battery
  • Over range Indication : "OL" will be shown on the LCD panel when out of
                                           range measurement is made
  • Operating Temperature and Humidity : 0¡É to 40¡É(32¢µ to 104¢µ), below 80% RH.
  • Storage Temperature and Humidity : -10¡É to 60¡É(14¢µ to 140¢µ), below 70% RH.
  • Accessories : Test leads Auxiliary¡æ1 Set, Auxiliary earth spikes¡æ2 Pcs,
                            Neck strap¡æ1 Pcs, Batteries 1.5V AA¡æ 8 Pcs, Manual